Recall Z-0012-2018 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Toshiba American Medical Systems Inc

Toshiba Medical Kalare Fluoroscopic X-Ray System (DREX-KL80) Kalare is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations. It is intended for use by qualified/trained doctors or technologists on both adult and pediatric subjects taking radiographic and fluoroscopic exposures of the whol

DateSep 12, 2017
StatusTerminated
ReasonDuring an examination images were displayed on the live monitor,but the images were not displayed on the system monitor nor were they saved to the hard disk.
Root causeSoftware design
ActionToshiba planned action to bring defect into compliance: 1. You will issue a Customer Notification Letter to make customers aware of the hazards, what measures will be taken to remedy the defect at no cost to the user, and how to safely use the equipment before corrections can be made. 2. You will install modified firmware on the systems to correct the defect and prevent recurrence. 3. You will provide progress reports on the effectiveness of the plan. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within
QuantityUS - 96
DistributionUSA (nationwide) Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133553KALARE
Toshiba America Medical Systems, In.C, Jun 13, 2014