Recall Z-0022-2018 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

St. Jude Medical, Inc.

Proclaim 5 Elite Implantable Pulse Generator and Proclaim 7 Elite Implantable Pulse Generator Models 3660, 3661, 3662, 3663, 3665, 3667

DateSep 12, 2017
StatusTerminated
ReasonAnalysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.
Root causeSoftware design
ActionSt. Jude Medical sent an Important Medical Device Advisory letter dated September 12, 2017, to all affected customers. Customers were advised the following: Phase I: ERI Assessment If a patient reports that an ERI message is displayed on the PC, please contact your Abbott Representative to obtain the device-specific generator logs and to conduct the ERI Assessment using the steps in Appendix B prior to making an elective replacement decision. After the ERI Assessment is completed, follow-up communication from your Abbott representative will be provided in order to make appropriate dec
Quantity28,817 devices
DistributionWorldwide Distribution - US (nationwide) (foreign consignee list will be provide by the firm at a later time)