Recall Z-0045-2018 ยท KQC Stains, Hematology
Sysmex America, Inc.
Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers
| Date | Oct 16, 2017 (initiated Sep 8, 2017) |
| Status | Completed |
| Reason | There is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I) result and the PLT-F result. In most cases, the falsely decreased PLT-F results display a PLT Abn Scattergram message with an asterisk beside the result or an Action Message: Difference between PLT and PLT- F. Check the results. These are indicators of unreliable data. In rare instances, platelet flags may be absent |
| Root cause | Nonconforming Material/Component |
| Action | Sysmex sent a Customer Letter dated August 2017 to all affected customers. The customer was directed to examine their inventory for the affected lot numbers and report back to Sysmex. .Customers with questions should call 1-888-879-7639, option 2 (if within the U.S.) or 1-888-679-7639 (for customers in Canada) |
| Quantity | 23,239 units (9353 US) |
| Distribution | Worldwide Distribution - US (nationwide including Puerto Rico) Internationally to New Zealand, Brazil, India, Philippines, Nicaragua, Argentina, Australia, Chile, Viet Nam, Colombia, Pakistan, Malaysia, Dubai, Germany, Hong Kong, Bangladesh, Uruguay Thailand, China, Saudi Arabia, Taiwan, Honduras, E |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.