Recall Z-0048-2018 ยท MIH System, Endovascular Graft, Aortic Aneurysm Treatment

W L Gore & Associates, Inc.

GORE TAG, THORACIC ENDOPROSTHESIS, CONFORMABLE. The GORE TAG Thoracic Endoprosthesis is intended for endovascular repair of all lesions of the descending thoracic aorta.

DateOct 25, 2017 (initiated Sep 25, 2017)
StatusTerminated
ReasonIncomplete and/or partial deployment of the CTAG Device during the endovascular procedure.
Root causeUnknown/Undetermined by firm
ActionThe firm contact their consignee with a safety notice by letter on 09/25/2017. Gore is updating its Instructions for Use (IFU) to include the following new warnings and precautions: " If abnormal or inconsistent deployment line resistance is felt during deployment initiation, STOP deployment action immediately. If device remains constrained, remove device through the introducer sheath. If resistance is felt during removal through the sheath, stop and withdraw device and introducer sheath together. " If the device is in a partially deployed state and remains attached to the catheter, phys
Quantity12,197units
DistributionWorldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.