Recall Z-0014-2018 ยท FSA Lift, Patient, Non-Ac-Powered

Handicare AB

Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009

DateJul 28, 2016
StatusTerminated
ReasonA limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.
Root causeProcess control
ActionThe firm initiated their recall by email on 07/28/2017. The firm directed the consignee to repair the units on site and destroy the affected parts.
Quantity9 units
DistributionOH, PA, MD
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.