Recall Z-0192-2018 ยท BYI Percussor, Powered-Electric

HILL-ROM MANUFACTURING, INC.

Monarch Airway Clearance System, Model 1000, Product Numbers: PMACS1NA (has blue tooth capability only) and PMACS1WI (has WIFI capability). The firm name on the label is Hill-Rom Services Pte. Ltd., Singapore. The Monarch Airway Clearance System, Model 1000, is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the thorax is the physicians

DateSep 15, 2017
StatusTerminated
ReasonMonarch Electromagnetic Interference (EMI) may disrupt the operation of implanted pacemakers or implantable cardioverter defibrillators as well as other devices that might be subject to an EMI if they come into direct contact with the Monarch device.
Root causeOther
ActionPhone calls were initiated September 15, 2017, to the customers, who are users. The expanded recall was initiated by letter dated November 13, 2017,via FedEx with delivery confirmation to both the users identified in the September notification and to the users who later received the recalled product. (Two local consignees will be provided the letter in person and one consignee will be sent the letter via USPS.) For further questions, please call (800) 426-4224.
Quantity60 devices
DistributionUSA (nationwide) Distribution to the states of: AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, MA, MI, MN, MO, NE, OH, OK, OR, TN, TX, VA, WA, WI, and WV. Internationally to Canada. There was no military/government distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163378Monarch Airway Clearance System
Hill-Rom Services Pte, Ltd., Mar 17, 2017