Recall Z-0021-2018 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

OriGen Biomedical, Inc.

The OriGen Reinforced Dual Lumen catheter (RDLC) is a duel lumen catheter with a wire reinforcement layer on the tip of the device. A thin sheath is used to cover and secure the wire layer and is heat shrunk into a permanent position.

DateOct 20, 2017 (initiated Aug 2, 2017)
StatusTerminated
ReasonTwo users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO Catheter has had a separation of the clear extension tube from the hub connection. These adverse events resulted in patient injury (blood loss).
Root causeUnder Investigation by firm
ActionThe firm initiated their recall by letter on 08/02/2017, following with a press release on 08/28/2017. The firm sent a revised letter to include instructions for the distributors on 09/22/2017 and an updated press release was sent on 09/26/2017. The firm requested that the consignee discontinue use of all OriGen VV28F Reinforced Dual Lumen ECMO Catheters from lots N18487 and N18487-1 and return any of those lots to OriGen Biomedical immediately for replacement or product credit.
Quantity180 units
DistributionUS Nationwide and the countries of Austria, Belgium, Chile, Czech Republic, Italy, Kingdom of Saudi Arabia, Netherlands, Poland, Sweden, Thailand, Colombia, India, South Africa, United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113869ORIGEN REINFORCED DUAL LUMEN CATHETER
Origen Biomedical, Inc., Jul 26, 2012