Recall Z-3239-2017 ยท FOX Stand, Infusion

Firefly Medical, Inc.

IVEA Product Part Number 16100 Model 500A

DateSep 7, 2017 (initiated Aug 24, 2017)
StatusTerminated
ReasonThe dowel pins (P/N 16082) that fit in the left and right leg pivot castings (P/Ns 16049/16056) can rotate and gradually maneuver out of the leg casting after several hundred deployment and folding cycles (designed for easy storage) of the IVEA unit.
Root causeDevice Design
ActionThe recall notification is expected to begin on or before 9/8/17 after pin cap components are received for the corrections. Instructions in the recall notification letter do not include return of recall product. They are to be corrected in the field. If any units are returned, they will be labeled and quarantined until corrected by Firefly Medical personnel or the sub-contracted manufacturer supplier. The recall notification letter will be sent certified letter with return receipt.
Quantity570 units
DistributionUnited States, Australia, Belgium, Canada, China, Germany, Hong Kong, Indonesia, Israel, Netherlands, Singapore, UAE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.