Recall Z-0046-2018 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.

DateAug 3, 2017
StatusTerminated
ReasonBeckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.
Root causeSoftware design
ActionBEC will initiate a SW update- to fix the SW defect. This will be executed via MOD (field modification) performed by Field Service.
Quantity27 units
DistributionChina France Italy Kuwait Spain Turkey United States