Recall Z-3178-2017 ยท HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment

Zimmer Biomet, Inc.

Zimmer Periarticular Locking Plate System, Locking Screw Tap Standard Quick-Connect 3.5 mm Diameter, Item Number/EDI 00236015335, Nonsterile. orthopedic surgical instrument.

DateSep 22, 2017 (initiated May 22, 2017)
StatusTerminated
ReasonVarious trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.
Root causeProcess control
ActionOn 5/22/2017, Zimmer Biomet distributed Urgent Medical Device Recall notices to their customers and distributors. On 12/19/2017, Zimmer Biomet sent Urgent Medical Device Recall notices to their customers due to a recall expansion involving one (1) lot for product # 20. Customers are advised to review the notice and ensure that affected personnel are aware of the contents. All stock should be inspected and product quarantined. A Zimmer Biomet sales representative will remove the affected product from your facility. Customers should complete and return the Certificate of Acknowledgement via
Quantity198 units
DistributionNationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.