Recall Z-3102-2017 ยท HTO Reamer

Encore Medical, Lp

MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58

DateAug 4, 2017
StatusTerminated
ReasonIt was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.
Root causeEmployee error
ActionThe firm initiated their recall by telephone and email to their sales representatives and surgeons on 08/04/2017. The letter explained the problem and advised the surgeons that the sales representative would be collecting the devices for return.
Quantity5 units
DistributionOH, VA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.