ClearanceBrief
Recall Z-3097-2017 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese patients. Procedures that can be performed with the AXIOM Artis-zee/zeego and Q/Q.zen family include cardiac angiography, neuro a

DateFeb 18, 2016
StatusTerminated
ReasonDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Root causeUnknown/Undetermined by firm
ActionSiemens mailed a Customer Safety Advisory Notice (CSAN) dated February 18, 2016, to affected customers informing them that their service organization will contact them to arrange a date to perform the corrective action (replacement of affected plastic couplings to prevent fluid leaks). Customers were advised to contact Siemens service organization for an appointment to schedule the work For questions regarding this recall call 610-.448-6461.
Quantity986 units in USA
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073290ARTIS ZEE AND ZEEGO
Siemens Medical Solutions USA, Inc., Feb 11, 2008