ClearanceBrief
Recall Z-3143-2017 ยท OLO Orthopedic Stereotaxic Instrument

Medtronic Navigation, Inc.

Spine Referencing Instrumentation, Set, Contains Spinous Process Clamps, Short and Tall When used with a Medtronic StealthStation Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable. The spine referencing clamps provide bony fixation t

DateJun 29, 2017
StatusTerminated
ReasonPotential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
Root causeDevice Design
ActionMedtronic sent an Urgent Medical Device Correction letter dated June 27, 2017, to all affected consignees. The letter notifies consignees of the product correction and requests that practitioners follow the attached use instructions for devices. Practitioners are to identify product damage before each procedure by opening and closing the device jaw with the T-handle. It is instructed that practitioners not continue turning the screw if resistance is met, as this may break the washer. Clamps are not to be used if the jaws remain closed when attempting to open a clamp or if the clamp screw back
Quantity2,051 units total
DistributionWorldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, AD, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, & WY. Foreign: Albania, Australia, Bangladesh, Belgium, Bulgaria, Canada, Croatia, Cz
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131425NAVIGATED CAPSTONE TRIALS, CLYDESDALE TRIALS, CAPSTONE & CLYDESDALE INSERTER
Medtronic Navigation, Inc., Aug 14, 2013