Recall Z-3113-2017 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc.

Coulter LH 750, Coulter LH 780, and Coulter LH 500 Analyzers Hematology Analyzers are quantitative, automated hematology analyzers and leukocyte differential counters For In Vitro Diagnostic used in clinical laboratories. They also provide automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fl

DateJun 26, 2017
StatusTerminated
ReasonBeckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.
Root causeLabeling Change Control
ActionBeckman Coulter sent an Urgent Medical Device Recall letter dated June 30, 2017, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to share the information with their laboratory staff. If they forwarded any of the affected product to another laboratory, please provide them a copy of the letter. Customers with questions should contact the Customer Support Center at 800-528-7694.
Quantity6763 units
DistributionWorldwide Distribution - US (nationwide) Internationally to Albania, Algeria, Andorra, Angola, Antigua and Barbuda, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Bulgaria, Burundi, Canada, China, Colombia, Congo, Croatia, Czech repu
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011342COULTER LH 750 HEMATOLOGY ANALYZER, MODEL 6605632
Beckman Coulter, Inc., Jul 9, 2001
K022161MODIFICATION TO COULTER LH 750 HEMATOLOGY ANALYZER, MODEL 6605632
Beckman Coulter, Inc., Jul 29, 2002
K030606COULTER LH 750 BODY FLUIDS APPLICATION
Beckman Coulter, Inc., Apr 21, 2003
K032342COULTER LH 750 HEMATOLOGY ANALYZER
Beckman Coulter, Inc., Sep 26, 2003
K050057COULTER LH 750 HEMATOLOGY ANALYZER
Beckman Coulter, Inc., Feb 10, 2005
K052905MODIFICATION TO COULTER LH 750 HEMATOLOGY ANALYZER
Beckman Coulter, Inc., Nov 18, 2005