Recall Z-0011-2018 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

ARCHITECT c4000 Cuvette Segment, List number 02P75-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL. Cuvette segments are a component to the ARCHITECT c4000 system. Cuvette segments are racks that sit in the reaction carousel and hold cuvettes. Each cuvette segment holds 11 cuvettes. With 9 segments, the reaction carousel holds 99 cuvettes. The Abbott ARCHITECT System is

DateJan 15, 2015
StatusTerminated
ReasonThere is a potential to generate falsely-depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000 instrument.
Root causeDevice Design
ActionThe firm, Abbott Laboratories, sent a "Product Correction" letter dated June 26, 2017 to is customers via overnight mail who had not had the Field Action Mandatory Technical Service Bulletin completed on their instrument. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: -Contact their representative to schedule an inspection of their cuvette segments. -If a detached segment is identified o A replacement segment will be provided at no charge. o Follow your laboratory protocol regarding previously reported patient results f
Quantity1,385 cuvette segments
DistributionWorldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belarus, Belgium, Bosnia & Herzegovina, Brazil, Brunei, Bulgaria Burkina Faso, Canada, Chile, China, Columbia, Costa Rica, Croati
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980367AEROSET SYSTEM
Abbott Laboratories, Apr 1, 1998