Recall Z-3039-2017 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)
Boston Scientific Corporation
SQ-RX 1010 Pulse Generator, Rx. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing
| Date | Aug 16, 2017 (initiated Jun 29, 2017) |
| Status | Terminated |
| Reason | The device can deliver an atypical amount of energy due to memory corruption inside the device. |
| Root cause | Software design |
| Action | The firm issued notifications dated June, 2017, beginning 6/29/2017 to physicians (implanting and patient follow-up physicians) via hand delivery by their sales representatives. The firm estimated approximately 30% of the U.S. physicians would be receiving the notification via hand delivery. Hand delivery by affiliates in foreign countries to customers began approximately 6/30/2017. Overnight mail letters were issued to the U.S. and foreign physicians starting on/about 7/7/2017 who did not receive hand- delivered notifications. |
| Quantity | Approximately 12,450 devices |
| Distribution | Worldwide - US Nationwide distribution, including Puerto Rico, U.S. Virgin Island, and Guam, There was also worldwide foreign distribution, including Canada. |
| FDA record | Open on fda.gov |