Recall Z-3042-2017 ยท DWJ System, Thermal Regulating

C.R. Bard, Inc.

Arctic Sun 5000 Temperature Management System, 100-120 V North America and 100-120 V US

DateAug 16, 2017 (initiated Mar 31, 2017)
StatusTerminated
ReasonPotential failure of the Arctic Sun 5000 Temperature Management System may occur when the continued use is in excess of 2000 system hours without preventive maintenance.
Root causeUnknown/Undetermined by firm
ActionBard sent an Urgent Customer Notification letter dated March 31, 2017, to each of the U.S. consignees. The letter contained recommended actions and included an Acknowledgement Form to be returned. For further questions, please call (770) 784-6119.
Quantity3,450 units
DistributionWorldwide Distribution - USA (nationwide) to the states of :AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WY., and to the countries of : Australia, Ko
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101092ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM, MODEL 5000
Medivance, Inc., Jul 16, 2010