ClearanceBrief
Recall Z-0007-2018 ยท NHA Abutment, Implant, Dental, Endosseous

Implant Direct Sybron Manufacturing, LLC

Implant Direct Swish Duo-Grip Healing Collar, Length: 3mm, Platform: 4.8 mm, Rx only, Sterile Dental Implant component.

DateJul 21, 2017
StatusTerminated
ReasonThe Swish Duo-Grip Healing Collar part number 9048-13H (4.8mm) package contained a 9037-13H (3.7 mm) component.
Root causeEmployee error
ActionRecall notification letters were sent to consignees, via mail on 07/21/2107. Instructions included to review current inventory and remove affected product, to complete and return the Acknowledgement and Recall Return form, and to notify customers who received product if it was further distributed.
Quantity50 units
DistributionDistribution US nationwide including CA, FL, NJ, NY, PA, TX, & UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090234SPECTRA SYSTEM DENTAL IMPLANTS 2008
Implant Direct, LLC, May 15, 2009