Recall Z-3073-2017 ยท FPD Tube, Feeding

Kentec Medical Inc

POLYURETHANE ENFit(R) AMERITUS NG FEEDING TUBE, REF Numbers: a) ENF-50P-40 b) ENF-50P-50 c) ENF-50P-80 d) ENF-60P-40 e) ENF-60P-50 f) ENF-60P-80 g) ENF-80P-40 h) ENF-80P-50 i) ENF-80P-80 j) ENF-100P-110 For pediatric and Neonatal Use, ENFit ENTERAL USE ONLY, RADIOPAQUE STRIPE, STERILE, EO, SINGLE USE ONLY, QTY 10/BOX.

DateJul 22, 2017 (initiated Jun 23, 2017)
StatusTerminated
ReasonInadequate detectability of radiopaque stripes
Root causeProcess control
ActionThe firm notified their consignees by telephone and email on 06/23/2017, and followed with a letter sent certified mail, on 07/07/2017. The letter instructed the consignee to cease use of the product, isolate it, and contact Kentec.
Quantity54 boxes of 10 individual units
DistributionTX, VA, IL, CA, KS