ClearanceBrief
Recall Z-2775-2017 ยท DWC Controller, Temperature, Cardiopulmonary Bypass

Sorin Group USA, Inc.

STOCKERT Heater-Cooler System 3T, 120V/60 Hz

DateJul 25, 2017 (initiated Jun 30, 2017)
StatusTerminated
ReasonPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
Root causeDevice Design
Action"Consignees were sent a recall notification, dated 06/30/2017, via mail. Instructions included a recommendation from LivaNova to use its oxygenators, or other oxygenators that are equipped with compatible heat exchangers, to ensure that the permeability rate is within the prescribed allowable limits (ALs). Depending on whether or not the oxygenator was from LivaNova, there were specific instructions to follow for each type and source. Consignees were also instructed to complete and return the Confirmation Form. If product was further distributed, the firm asks that further customers b
Quantity7,546 units
Distribution"Domestic: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, & WY. Foreign: AL, AE, AR, AT, AU, AZ, BD, BE, BG, BH, BR, BY, CA, CH, CL, CN, CO, CR, C
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052601STOCKERT HEATER-COOLER SYSTEM 3T
Sorin Group Deutschland GmbH, Jun 6, 2006