Recall Z-3040-2017 ยท FKX System, Peritoneal, Automatic Delivery
Baxter Healthcare Corporation
HomeChoice Automated PD system Model numbers: 5C4471, 5C4471R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.
| Date | Jul 5, 2017 |
| Status | Terminated |
| Reason | Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient. |
| Root cause | Use error |
| Action | The firm notified their consignees by letter on 07/05/2017 of the potential for damaging the disposable set if tools are used to open the packaging and directing them to only open the packaging by hand by letter. |
| Quantity | 15144 units |
| Distribution | Worldwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K102936 | HOMECHOICE/HOMECHOICE PRO AUTOMATED PERSONAL CYCLER PERITONEAL DIALYSIS SYSTEM MODEL 5C4471, 5C8310, 5C4471R AND 5C8310R Baxter Healthcare Corp, Mar 30, 2011 |