Recall Z-3035-2017 ยท DOF Lead, Atomic Absorption

Magellan Diagnostics, Inc.

LeadCare II Blood Lead Test Kit, containing LeadCare II Lead Controls Product Usage: IVD test kit for the detection of lead in blood. The kit consists of sensors, treatment reagent, lead controls (two levels), heparinized capillary tubes and plungers, transfer droppers and a calibration button. The LeadCare II Lead Controls packaged in the kit lots listed in #2 below are the subject of this re

DateAug 15, 2017 (initiated Jan 20, 2016)
StatusTerminated
ReasonLeadCare II Test Kits controls out of range
Root causeDevice Design
ActionMagellan notified customers initially by Email, fax and/or verbal communication with Magellan Product Support representatives beginning on January 20, 2016 to advise of the problem and provide revised Assignment Control Value Matrix..
Quantity12,931 kits
DistributionWorldwide - US Nationwide distribution and the countries of: COLOMBIA, CANADA, THAILAND, INDIA, UAE, AUSTRALIA, GERMANY, CANADA, SOUTH AFRICA, SPAIN, NEW ZEALAND, KOREA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052549LEADCARE II BLOOD LEAD TESTING SYSTEM
Esa Biosciences, Inc., Oct 6, 2005