Recall Z-0178-2018 ยท HAW Neurological Stereotaxic Instrument

Zimmer Biomet, Inc.

ROSA Surgical Device 2.5.8

DateFeb 10, 2013
StatusTerminated
ReasonIssue with optional neuro-endoscopy module detected under specific conditions during internal testing.
Root causeSoftware design
ActionField Service Technicians were dispatched to correct units by upgrading the system to ROSA 2.5.7. A further update of ROSA 2.5.8.4. A Field Safety Notice dated 07/10/2013 notifying consignees of a risk in the ROSA system recognized in internal testing. It was suggested that consignees not utilize the neuro-endoscopy module until further notice. The recall notice should have been shared with the appropriate parties and the signed acknowledgement form returned to the recalling firm.
Quantity22 units (4 US and 18 OUS)
DistributionAR, OH, MI, and TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092239ROSA SURGICAL DEVICE, MODEL ROSA 1.1
Medtech S.A, Nov 17, 2009
K101791ROSA SURGICAL DEVICE
Medtech S.A.S, Sep 23, 2010