Recall Z-2866-2017 ยท GEX Powered Laser Surgical Instrument

Xintec Corp

OptiLITE(TM) Products for Laser Surgery, Catalog Numbers: HTM1020F, HTM1040F, HTM1060F, HTM1010F; Manufactured by CONVERGENT LASTER TECHNOLOGIES The Odyssey 30B is approved for numerous medical procedures. Generally, the primary use of the laser (which includes the fibers) is related to urology, but ultimately it is up to the surgeon except in cases where it is contraindicated. The fibers are

DateJun 23, 2017
StatusTerminated
ReasonIt has been determined that the reprocessing instructions may not provide sufficiently detailed information for the cleaning, disinfection and sterilization of the products listed below. The preliminary investigation indicates that validation data related to the reprocessing of these devices may not meet newly enacted standards.
Root causeLabeling False and Misleading
ActionThe firm, Convergent Laser Technologies, sent an "URGENT - MEDICAL DEVICE RECALL" letter dated 06/23/2017 to its Consignees. The letter described the product, problem and actions to be taken. The consignees were instructed as follows: "Please immediately review your inventory and quarantine any affected product. Please return any affected products within 30 days according to the instructions provided in the attached Acknowledgement and Receipt Form. Upon receipt of the product and response form, we will arrange for a credit to your account. Also, whether or not you are returning any affect
Quantity4343 units
DistributionWorldwide distribution- US (Nationwide) and countries of: Canada and Sri Lanka.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992866OPTILITE IX LASER SURGERY ACCESSORIES
Convergent Laser Technologies, Nov 18, 1999