ClearanceBrief
Recall Z-2764-2017 ยท DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass

Medtronic Perfusion Systems

DLP Extension Line Adapters 20 in length Used to extend cardioplegia line during cardiopulmonary bypass surgery

DateJul 10, 2017 (initiated Jun 22, 2017)
StatusTerminated
ReasonIdentification of small pinholes in a single packaging configuration of sterile pouches. A total of seven different products were affected.
Root causePackaging process control
ActionMedtronic sent an Urgent Medical Device Recall letter dated June 22, 2017, to all affected consignees, both in the US, via 2-Day UPS Mail, and OUS, via local methods. Consignees were asked to quarantine and return any affected product. No recommendations were made for patients who have undergone procedures using affected product. US consignees were asked to complete and return a recall notification confirmation certificate to fax 651-367-0612, indicating that they were informed of the recall and executed the appropriate action. Customers with questions were instructed to contact Customer Serv
Quantity63053 (48180 US) (14873 OUS)
DistributionWorldwide Distribution - US (nationwide) Foreign: Argentina, Australia, Austria, Belgium, Bulgaria, Canada, Croatia, Denmark, El Salvador, France, Georgia, Germany, Hungary, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Macedonia, Moldova, Netherlands, New Zealand, Nigeria, Norway, Poland, Romania
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K790564ADAPTER-STRAIGHT
Dlp, Inc., Apr 12, 1979