Recall Z-0530-2018 ยท NEY System, Ablation, Microwave And Accessories

Angiodynamics, Inc.

Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601-US, Model # (UPN) H787900601US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

DateJun 5, 2017
StatusTerminated
ReasonThe firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.
Root causeUnder Investigation by firm
ActionURGENT VOLUNTARY MEDICAL DEVICE RECALL Letters and Response Forms (dated June 5, 2107) were sent to consignees via FedEx on June 5, 2017. The letter identified the affected product, problem and actions to be taken.
Quantity14,353 units in total
DistributionWorldwide distribution - US Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K094021ACCULIS ACCU2I PMTA APPLICATOR MODEL ACCU2I
Microsulis Medical Limited, Aug 6, 2010