Recall Z-0160-2018 ยท HAW Neurological Stereotaxic Instrument

Zimmer Biomet, Inc.

ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments.

DateNov 6, 2014
StatusTerminated
ReasonPotential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).
Root causeDevice Design
ActionAll 9 domestic units were corrected by the firm in February 2017, via technician visits to consignee sites.
Quantity31 units
DistributionUS distribution in AR, GA, OH, MI, and TX.
FDA recordOpen on fda.gov