ClearanceBrief
Recall Z-3031-2017 ยท FRN Pump, Infusion

ICU Medical Inc

Plum 360 Infusion System

DateJun 15, 2017
StatusTerminated
ReasonPotential for the Connectivity Engine Module of the affected unit to disengage from the main chassis, causing the unit to shut down without an alarm notification. This would delay initiation of therapy or interrupt an infusion.
Root causeUnder Investigation by firm
ActionICU Medical sent an Urgent - Medical Device Correction letter dated June 15, 2017, to all affected customers. The firm contracted with Stericycle for recall notification, response tracking, and effectiveness checks. The customer notification was also posted on icumed.com. The notification requested that consignees verify their unit is working and testing the CE Module. If the unit's CE Module is loose or a blank display occurs, customers are asked to record the S/N and contact ICU Medical Technical Service Support Center at 800-241-4002, option 4. If the unit is functional, return to servi
Quantity68,559 units
DistributionWorldwide Distribution - USA (nationwide) Distribution to the states of including Puerto Rico : AL, AR, AZ, CA, FL, GA, HI, IA, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NE, NH, NJ, NM, NY, OH, PA, SD, TN, TX, VA, WA, WI, WY and District of Columbia., and to the countries of : Australia, Cana
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141789PLUM 360 INFUSION SYSTEM WITH HOSPIRA MEDNET, SMART CARD PLUG N PLAY CE 3.0 MODULE FOR PLUM A+ INFUSION SYSTEM UPGRADE
Hospira, Inc., Jan 16, 2015
K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play Module
Hospira, Inc., Mar 28, 2017