ClearanceBrief
Recall Z-2712-2017 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Toshiba American Medical Systems Inc

APLIO i-series: a) APLIO i700 TUS AI700 b) APLIO i800 TUS AI800 diagnostic ultrasound system

DateMay 1, 2017
StatusTerminated
ReasonWhen Continuous Trace method is used as the tracing method on Spectral Doppler in velocity trace measurement, an incorrect value may be displayed in the ultrasound systems and in the ultrasound workstations listed below.
Root causeSoftware Manufacturing/Software Deployment
ActionToshiba Medical sent an Urgent Medical Device Correction letter on May 1, 2017. The consignee is instructed not to use the Continuous Trace method for cardiac velocity trace measurements to quantify the MPG values. Modified software is being provided. A Toshiba service representative will contact the consignee to make an appointment to install the modified software to fix the problem. Customers were asked to complete the form and fax it to the toll-free number shown at the top of the form. The form can also be sent by email to RAffairs@tams.com. Customers with questions were instructed
Quantity28 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161843Aplio i900/i800/i700 Diagnostic Ultrasound System, V2.0
Toshibamedical Systems Corporation, Sep 21, 2016