ClearanceBrief
Recall Z-2710-2017 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Toshiba American Medical Systems Inc

APLIO 300 TUS-A300 diagnostic ultrasound system

DateMay 1, 2017
StatusTerminated
ReasonWhen Continuous Trace method is used as the tracing method on Spectral Doppler in velocity trace measurement, an incorrect value may be displayed in the ultrasound systems and in the ultrasound workstations listed below.
Root causeSoftware Manufacturing/Software Deployment
ActionToshiba Medical sent an Urgent Medical Device Correction letter on May 1, 2017. The consignee is instructed not to use the Continuous Trace method for cardiac velocity trace measurements to quantify the MPG values. Modified software is being provided. A Toshiba service representative will contact the consignee to make an appointment to install the modified software to fix the problem. Customers were asked to complete the form and fax it to the toll-free number shown at the top of the form. The form can also be sent by email to RAffairs@tams.com. Customers with questions were instructed
Quantity90 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151451Aplio 500/400/300 Diagnostic Ultrasound System V6.0
Toshibamedical Systems Corporation, Jul 9, 2015