Recall Z-0180-2018 ยท LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Arrow International Inc
ARROW PICC powered by ARROW VPS Stylet The Pressure Injectable PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, power injection of contrast media and allows for central venous pressure monitoring
| Date | Jun 1, 2017 |
| Status | Terminated |
| Reason | There is a possibility that the finished good kits may contain 5 Fr peel-away sheaths instead of the required 4 Fr peel-away sheaths. |
| Root cause | Mixed-up of materials/components |
| Action | Arrow International sent an Urgent Medical Device Recall Notification letter dated June 1, 2017, and Recall Acknowledgement Form to their domestic consignees via FedEx 2-day. Customers were advised to immediately inspect stock, discontinue use and quarantine any products with the product codes and lot numbers listed, so that the affected products can be returned to Arrow. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow the firm to document the amount of product you hav |
| Distribution | Nationwide Distribution to AZ, CA, MA, MO, NC, NY, OR & SD |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K121941 | ARROW PICC POWERED BY ARROW VPS STYLET Arrow International (Subsidiary of Teleflex Inc.), Sep 7, 2012 |