ClearanceBrief
Recall Z-0210-2018 ยท LPB Cardiac Ablation Percutaneous Catheter

Biosense Webster, Inc.

SMARTABLATE RF System Generator

DateJun 14, 2017
StatusTerminated
ReasonGenerator may inadvertently deliver RF energy if the foot pedal is placed in close proximity and oriented toward the Stereotaxis magnets.
Root causeLabeling design
ActionAll new U.S. and OUS customers, who have purchased the SMARTABLATE" Generator system since the launch of the original field action in March 2015 to the time the IFU was updated in November 2015, will be notified. The previously distributed Field Safety Notification Letter for the SMARTABLATE" Generator (Part Numbers# M490006 and M490007) and customer Acknowledgement Form will be sent to these new customers. Instructions included review the Field Safety Notice carefully, provide the notice to appropriate personnel in facility, ensure the SMARTABLATE System Foot Pedal is not located within th
Quantity185 units total
DistributionDistributed to the following states: CA, CO, GA, IL, IN, MI, NM, NY, OH, PA, TN, TX, UT, WA, WI. Product was also distributed to the following countries: Germany, France, Austria, Sweden Italy, Belgium, Denmark, and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.