Recall Z-2702-2017 ยท NRY Catheter, Thrombus Retriever

Penumbra Inc.

Penumbra 3D Revascularization Device It is indicated for used in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral-M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are ca

DateJun 22, 2017 (initiated Jun 9, 2017)
StatusTerminated
ReasonPenumbra has identified an issue in these four lots involving a raw material component of the delivery wire. This issue could result in breakage of the delivery wire, which could potentially lead to serious patient injury or death.
Root causeComponent change control
ActionPenumbra sent an Urgent Voluntary Field Removal Notice dated June 9, 2017. Customers were instructed to inspect current inventory and remove any affected units for return. Customers were also instructed to return the response form. Penumbra has identified an issue in these four lots involving a raw material component of the delivery wire. This issue could result in breakage of the delivery wire, which could potentially lead to serious patient injury or death. Penumbra personnel will contact customers directly to arrange return of affected units and replacement of returned product at n
Quantity155 units
DistributionNationwide Distribution to the following states: AZ, CA, CO, FL, IN, MA, MI, MN, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162901Penumbra 3D Revascularization Device
Penumbra, Inc., Apr 20, 2017