ClearanceBrief
Recall Z-2738-2017 ยท DSP System, Balloon, Intra-Aortic And Control

Maquet Datascope Corp - Cardiac Assist Division

CS 300 Intra-Aortic Balloon Pump

DateJun 23, 2017 (initiated Jun 16, 2017)
StatusTerminated
ReasonThe device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.
Root causeComponent design/selection
ActionDatascope Corp/MAQUET issued a press release on June 16, 2017. Datascope Corp/MAQUET sent an Urgent Product Recall Letter on June 19, 2017, to their customers via FedEx next day delivery service. Pursuant to the WARNINGS section of our CS100i, CS100 or CS300 IABP Operating/User Instructions, clinicians are instructed not to leave the patient unattended during IABP therapy. Until the service is performed, we recommend the IABP be powered on prior to initiation of therapy to allow the IABP to successfully complete its self-test. This action will take less than 60 seconds to perform. In the e
Quantity12,319 units total
DistributionWorldwide Distribution - (in over 100 countries) US (Nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151254CARDIOSAVE Hybrid Intra-Aortic Balloon Pump
Datascope Corp., Jul 2, 2015