Recall Z-0191-2018 ยท CHN Immunochemical, Ceruloplasmin

CSL Behring GmbH

SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/Product Code OQIN19, SMN 10446079 (Low); N/T Protein Control SL/M, REF/Product Code OQIO19, SMN 10446085 (Mid); N/T Protein Control SL/H, REF/Product Code OQIP19, SMN 10446089 (High);

DateJun 5, 2017
StatusTerminated
ReasonSiemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding
Root causeNonconforming Material/Component
ActionURGENT MEDICAL DEVICE CORRECTION Letter (PP17-011.a.US, dated May 30, 2017) was sent to US consignees via FedEx on 6/5/2017. URGENT FIELD CORRECTIVE ACTION Letter was e-mailed to customers Outside the US. Customers are instructed to use the re-assigned Ceruloplasmin values for the lots indicated in the letter.
DistributionWorldwide including the countries of United States, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia-Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Great Britain, Hungary, India, I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052788N LATEX HCY, N PROTEIN STANDARD SL, N/T PROTEIN CONTROL L/M/H
Dade Behring, Inc., Mar 29, 2006