Recall Z-2691-2017 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
St. Jude Medical, Inc.
Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electri
| Date | Jun 2, 2017 |
| Status | Terminated |
| Reason | The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures. |
| Root cause | Device Design |
| Action | The notification material is in the form of a Physician Letter and Patient letter which will be delivered to physicians and patients in the US. The physician mailing will be initiated June 2, 2017 for delivery on June 7-9, 2017. The physicians were provided with patient management recommendations, Step for the patient controller updates, and Clinician's Manual Revised Warning. The Patient mailings will begin the week of June 12, 2017 to allow for clinician notification and are expected to be delivered within 3-5 business days. The patient letter informs of them of the potential for the l |
| Quantity | 8780 units |
| Distribution | US Nationwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.