ClearanceBrief
Recall Z-3029-2017 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Medical Systems, LLC

Signa PET/MR (K163619, K142098) Product Usage: The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically. The combined system maintains independent functionality of the MR and PET devices, al

DateAug 2, 2017 (initiated May 11, 2017)
StatusTerminated
ReasonPossible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Root causeMixed-up of materials/components
ActionConsignees received a written recall notice dated 05/17/2017 notifying them of a correction to all units using an Uninterruptable Power Supply. Consignees were instructed not to unplug the unit and to continue normal use. Units with the Eaton 9130 Uninterruptable Power Supply connected to the MR operator console would receive a correction at no charge. This would involve possibly replacing the power cable. GE Representatives will be contacting consignees to arrange corrections. Any questions can be directed to GE Healthcare Service at 1-800-437-1171 or your local service representative.
Quantity68 units in total
DistributionUS Nationwide in the states of : AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142098SIGNA PET/MR
Ge Medical Systems, LLC, Nov 24, 2014
K163619SIGNA PET/MR
Ge Medical Systems, LLC, Feb 14, 2017