Recall Z-2621-2017 ยท IWE Monitor, Patient Position, Light-Beam

Omega Medical Imaging, Inc.

Omega Medical Imaging - 2 over 2 Elevating Monitor Suspension pillar/column accessory Device accessory that elevates monitor for healthcare professionals.

DateJun 19, 2017 (initiated Mar 23, 2017)
StatusTerminated
ReasonElevating Monitor Suspension column dropping and will not rise back up.
Root causeMixed-up of materials/components
ActionThe firm, OMEGA MEDICAL IMAGING, sent a "Field Safety Advisory Notice" dated 3/23/2017 to consignees. The notice described the product, problem and actions to be taken. The consignees were instructed to immediately examine the product and confirm correct product information; immediately discontinue use; move monitor suspension away from staff personnel and patients; and advising consignees to take photos of affected product and the P/N label. Photos of product labels are to be sent to jnewman@omegamedicalimaging.com at Omega. Product is to be corrected. If there are any questions or concer
Quantity6 units
DistributionWorldwide Distribution -US Distribution to states of: IN, OH, PA, and MT; and country of: Pakistan.