Recall Z-2867-2017 ยท JFY Enzymatic Method, Creatinine

Sentinel CH SPA

MULTIGENT Creatinine (Enzymatic), Catalog Numbers 8L24-31 and 8L24-41; IVD. Intended for the quantitative determination of creatinine in human serum, plasma, and urine using the ARCHITECT c Systems The product is a medical device (chemistry reagent) and is intended for the quantitative determination of creatinine in human serum, plasma, and urine using the ARCHITECT c Systems. The kit is provide

DateJun 27, 2017 (initiated May 22, 2017)
StatusTerminated
ReasonReagent 1 component may result in Quality Control values shifting out of acceptable range, which is not resolved by recalibration.
Root causeNonconforming Material/Component
ActionSentinel Diagnostics notified the US Distributor on 5/22/17 to notify US customers by letter. Actions to be Taken 1. Discontinue the use of the impacted lots immediately and destroy any remaining inventory of impacted lots according to your laboratory procedures. 2. If you have an alternate reagent lot available in inventory, switch to this alternate lot, or use a new lot provided. 3. Please review the content of this communication with your Medical Director and retain this letter for any future reference. Questions regarding this information, U.S. Customers should contact Customer Servic
Quantity2356 kits Worldwide; US 206 kits
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073634MULTIGENT CREATININE (ENZYMATIC) ASSAY
SENTINEL CH. SpA, Jun 19, 2008