Recall Z-0173-2018 ยท CHN Immunochemical, Ceruloplasmin

Siemens Healthcare Diagnostics, Inc.

Dimension Vista Protein 1 Calibrator, PROT 1 CAL, Siemens Material Number (SMN) 10465664 (US) and 10469516 (OUS), Product Code KC710U (US) and KC710 (OUS); IVD

DateMay 23, 2017
StatusTerminated
ReasonComplaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).
Root causeNonconforming Material/Component
ActionSiemens issued an Urgent Medical Device Correction (UMDC) letter (PP-17-012.A.US, dated May 23, 2017) to US customers on May 23, 2017 and an Urgent Field Safety Notice (UFSN) to customers Outside the US in May 2017. Customers were instructed in the letter to use re-assigned CER values for the lots indicated of PROT1 CAL and PROT1 CON L, M, and H. In the US, the method of distribution is FEDEX and outside the US, the distribution is determined at the country level. Customers were requested to complete Effectiveness Check questionnaire attached to the UMDC/UFSN and return to Siemens within 30
DistributionWorldwide Distribution - US (nationwide) Internationally to Austria, Australia, Belgium, Canada, France, Germany, Great Britain, Italy, Japan, Netherlands, Portugal, Slovakia, Spain, Switzerland and Republic of Korea South
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082503DIMENSION VISTA LG LIGHT CHAINS, TYPE KAPPA, DIMENSION VISTA LG LIGHT CHAINS, TYPE LAMBDA, DIMENSION VISTA PROTEIN 1 CAL
Siemens Healthcare Diagnostics, Dec 19, 2008