ClearanceBrief
Recall Z-3067-2017 ยท ODG Endoscopic Ultrasound System, Gastroenterology-Urology

Pentax of America Inc

Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.

DateMay 23, 2017
StatusTerminated
ReasonError in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
Root causeError in labeling
ActionRecall Notifications were sent to 104 consignees via USPS Certified Mail on 5/23/2017. Included with the Field Safety letter was the customer response form, and the addendum to IFU Z845. These were also posted on Pentax of America web page until the recall is closed.
QuantityUS: 118 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061068EG-3670URK
Pentax Medical Company, May 3, 2006
K090196EG-3670URK
Pentax Medical Company, May 19, 2009
K130706DUOFLEX COIL SUITE (1.5T)
Mr Instruments, Inc., Jul 25, 2013