Recall Z-3067-2017 ยท ODG Endoscopic Ultrasound System, Gastroenterology-Urology
Pentax of America Inc
Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.
| Date | May 23, 2017 |
| Status | Terminated |
| Reason | Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step. |
| Root cause | Error in labeling |
| Action | Recall Notifications were sent to 104 consignees via USPS Certified Mail on 5/23/2017. Included with the Field Safety letter was the customer response form, and the addendum to IFU Z845. These were also posted on Pentax of America web page until the recall is closed. |
| Quantity | US: 118 units |
| Distribution | Nationwide |