ClearanceBrief
Recall Z-2191-2017 ยท KPS System, Tomography, Computed, Emission

GE Healthcare, LLC

GE Healthcare Discovery Ml DR X-ray system

DateMay 19, 2017
StatusTerminated
ReasonIt was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.
Root causeSoftware design
ActionGE Healthcare proposed action to bring defect into compliance : 1. Inform users of the potential for this issue to occur and instruct users as to the actions to take in the event of an Emergency Stop actuation during prep-delay. 2. Provide an update to product labeling for users that includes this information. GE Healthcare will include the updated product labeling with all current production systems. 3. Make modifications during the next scheduled product update to prevent this "race" condition from occurring in newly produced systems. CDRH approves the CAP subject to the followi
Quantity10,884
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Japan, Jordan, Kazakhstan, Korea ( Republic of) Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritania, Mexico, Moldova, Morocco, Nambia, Netherlands, New Caledonia, New
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133657DISCOVERY PET/CT 710
Ge Medical Systems, LLC, Mar 21, 2014