ClearanceBrief
Recall Z-2689-2017 ยท PER Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis

Beckman Coulter Inc.

Coulter PrepPlus, Part Number 286600, Product Usage: The COULTER PrepPlus is a microprocessor-controlled pipetting and diluting system designed for automating sample preparation or assay methods. It is capable of aspirating and dispensing liquid samples.

DateMay 22, 2017
StatusTerminated
ReasonWhen samples are prepared using the PrepPlus or PrepPlus 2 sample preparation systems and then run on a Flow cytometer, there is an imprecision in the absolute counts on only the first tube in a run.
Root causeSoftware design
ActionThe recalling firm initiated their recall by letter disseminated via email on 05/22/2017. The letter requested the following actions be taken: 1. Insert an empty daughter tube in place of a patient sample as the first tube in the carousel that is programmed for the addition of Flow-Count Fluorospheres. 2. Refer to the COULTER PrepPlus Operators Guide / PrepPlus 2 Instructions for Use (IFU) for instructions on adding Flow-Count Fluorospheres to daughter tubes. 3. Review all results following the flow cytometer data review instructions referenced in the associated IFU. Beckman Coul
Quantity43 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130253TQ-PREP WORKSTATION, PREPPLUS 2 WORKSTATION
Beckman Coulter, Inc., Jul 15, 2013