Recall Z-2655-2017 ยท KPR System, X-Ray, Stationary

Shimadzu Medical Systems Usa Com

SHIMADZU FDR Visionary Catalog Number: FDR Visionary Suite Stationary x-ray system for taking general radiographic and tomographic images of the whole body. The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals

DateMay 23, 2017
StatusTerminated
ReasonCracks may occur over time on the tube holding shaft near the base of the tube mounting flange.
Root causeDevice Design
ActionShimadzu sent an Urgent Voluntary Medical Device Recall Notice to all affected customers. The firm will implement a voluntary recall/field corrective action to add protective parts preventing the x-ray tube assembly from coming into contact with the patient or operator should separation of the tube mounting flange from the tube holding shaft occur. Customers with questions were instructed to call 800-228-1429.
Quantity1673 units total
DistributionWorldwide Distribution - Nationwide distribution to AK, AR, AZ, CA, CT, FL, IL, IN, KS, LA, MI, MO, NE, NJ, NM, NV, NY,OH, OK, PA, RI, SC, TN, TX, WA, WI, and Hawaii Foreign: Canada and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152294FDR Visionary Suite
Shimadzu Corporation, Jul 15, 2016