Recall Z-2653-2017 ยท KPR System, X-Ray, Stationary

Shimadzu Medical Systems Usa Com

SHIMADZU Ceiling Type X-ray Tube Support CH-200/CH-200M Catalog Number: CH-200/CH-200M Holds an x-ray tube unit and collimator combined with an z-ray high voltage generator, radiographic stand, and table. It is designed to perform radiography of patients either in a standing or supine position

DateMay 23, 2017
StatusTerminated
ReasonCracks may occur over time on the tube holding shaft near the base of the tube mounting flange.
Root causeDevice Design
ActionShimadzu sent an Urgent Voluntary Medical Device Recall Notice to all affected customers. The firm will implement a voluntary recall/field corrective action to add protective parts preventing the x-ray tube assembly from coming into contact with the patient or operator should separation of the tube mounting flange from the tube holding shaft occur. Customers with questions were instructed to call 800-228-1429.
Quantity1673 units total
DistributionWorldwide Distribution - Nationwide distribution to AK, AR, AZ, CA, CT, FL, IL, IN, KS, LA, MI, MO, NE, NJ, NM, NV, NY,OH, OK, PA, RI, SC, TN, TX, WA, WI, and Hawaii Foreign: Canada and Japan