ClearanceBrief
Recall Z-2324-2017 ยท EOQ Bronchoscope (Flexible Or Rigid)

Pentax of America Inc

Video Bronchoscope

DateJan 11, 2011
StatusTerminated
ReasonThis field action serves to retrospectively document the actions that were taken by Pentax to correct the loose suction arms on eleven (11) models of endoscopes and to notify FDA of these actions taken in 2010-2011.
Root causeDevice Design
ActionCustomers were visited by members of Pentax America, Inc. field service team. The field service technicians inspected the affected devices, and documented the inspections using the Inspection Record Form (QS-397).
Quantity1,349 - (Total - US) and 352 (Total - OUS)
DistributionUS Nationwide and Internationally
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131028PENTAX VIDEO BRONCHOSCOPES (EB FAMILY)
Pentax Medical Company, Apr 10, 2014
K131946ENTAX ULTRASOUND VIDEO BRONCHOSCOPE EB-1970UK + HI VISIN PREIRUS
Pentax Medical Company, Apr 25, 2014