Recall Z-2713-2017 ยท FDT Duodenoscope And Accessories, Flexible/Rigid

Pentax of America Inc

Video Duodenoscope These instruments are intended to provide optical visualization of (via a video monitor), and therapeutic access to, the biliary tract via the upper GI tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: esophagus, stomach, duodenum, common bile, hepatic and cystic ducts.

DateJan 17, 2017
StatusTerminated
ReasonPentax initiated a field correction/safety alert for two (2) models of the Video Duodenoscope to determine how soiling may have occurred on the surface of the suction cylinder and under the distal cap during testing.
Root causeDevice Design
ActionThe firm, Pentax Medical, sent a "FIELD CORRECTION' letter dated January 17, 2017 to their affected customers via USPS Certified Mail. The letter described the product, problem and actions to be taken. The customers were instructed to immediately remove any affected product from use; follow product labeling; ensure all reprocessing personnel are knowledgeable and thoroughly trained on the instructions of Use for manual reprocessing of the devices; clean elevator recesses and follow all reprocessing instructions; and complete and return the FIELD CORRECTION RESPONSE FORM via a member of the PEN
Quantity2,015 (US - 519 (ED-3490TK) and 5 (ED-3270K); OUS - 1,491)
DistributionWorldwide Distribution-US (Nationwide) including states of: AL, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, PR, TN, TX, UT, VA, WA, WI, WV, WY;. and country of: CANADA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092710PENTAX ED-3490TK, VIDEO DUODENOSCOPE
Pentax Medical Company, Dec 2, 2009
K963056ED-3410, VIDEO DUODENOSCOPE
Pentax Precision Instrument Corp., Oct 10, 1996