ClearanceBrief
Recall Z-2278-2017 ยท FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

Arrow International Inc

AGB+ Pressure Injectable Multi-Lumen CVC Kit intended to permit venous access to the central circulation by way of the femoral, jugular or subclavian veins intended to permit venous access to the central circulation by way of the femoral, jugular or subclavian veins

DateMar 20, 2017
StatusTerminated
ReasonArrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.
Root causeUnder Investigation by firm
ActionArrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002507TWO-LUMEN CENTRAL VENOUS ACCESS KIT WITH HEMOSTASIS VALVE/SIDE PORT
Arrow Intl., Inc., Mar 21, 2001