Recall Z-2323-2017 ยท FMT Warmer, Infant Radiant

GE Healthcare, LLC

Infant Warmer System (IWS)

DateMay 19, 2017 (initiated May 5, 2017)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.
Root causeUnder Investigation by firm
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated May 5, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Product Correction : Attached to this letter, we provide instructions as part of a Service Manual Addendum on how to correct the issue. Please add this new Addendum to the Service Manual of your device(s) and train the affected users accordingly. Please acknowledge that you have received this letter and that you understand that an action needs to be taken on your part to correct th
Quantity34986
DistributionWorldwide Distribution - USA (nationwide) Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K840858OHIO 3300 INFANT WARMER SYSTEM
Ohio Medical Products, Mar 23, 1984
K851449OHMEDA 5000 INFANT WARMER SYSTEM
Ohmeda Medical, Jun 7, 1985
K884160OHMEDA - OHIO(R) 3500 INFANT WARMER SYSTEM
Ohmeda Medical, Nov 21, 1988
K921766OHIO INFANT WARMER SYSTEM
Ohmeda Medical, Apr 23, 1993
K963058INFANT WARMER SYSTEM
Ohmeda Medical, Oct 1, 1996