Recall Z-2252-2017 ยท GDR Saw, Manual And Accessories

Argon Medical Devices, Inc

Clear Flashback Needle 7cm x 18G AMC/4; Clear Flashback Needle 2.5 cm x 21G AMC/4; Winged Arterial Needle 7cm x 18G AMC/4; Clear Flashback Needle 7 cm x 19G AMC/4 Usage: Argon Arterial Access Needles are designed to provide vascular access for the percutaneous introduction of a guidewire.

DateMay 11, 2017
StatusTerminated
ReasonArgon Medical has received a complaint from one of their distributors of a potential packaging defect. The product is sealed in a pouch and the defect if present shows as a potential void or unsealed edge of the pouch.
Root causeProcess design
ActionArgon Medical Devices, Inc. is sending an Urgent Medical Device Recall letter by registered mail to the affected customer/hospital/distributor. Customers are instructed to inspect and quarantine and return any affected product remaining at their facility utilizing the instruction on the Recall Response Form.
Quantity7140 pieces
DistributionWorldwide Distribution - US Nationwide in the countries of Belgium, Canada, Japan, Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.